Exos Leak/O2/CO2

6 tests in 1: O2/CO2 gas analysis; micro-leak measurement system; creep test (A&B); packaging sealing resistance

The EXOS LEAK/O2/CO2 allows you to measure the rate of O2/CO2 and the airtightness for all types of packaging, containers and pneumatic circuits. This measurement device also checks the resistance of sealing using a bursting test. These three tests can be carried out in quick succession on the same packaging. The EXOS lets you inspect the integrity of your packaging, and can be used in the laboratory as well as during the manufacturing process.

THE + PRODUCTS

  • Meets standards:
    DIN 55508-1: measures flow rate at constant pressure
    ASTM F1140: seal strength
    ASTM F2054: burst test
  • The option of combining 6 measurements in 1 same test: O2/CO2 gas analysis; Micro-leaks; Creep test (A&B); Bursting
  • Blocked circuit detection
  • Measures with low excessive pressure
  • Breathability measurement (micro-perforated films)
  • Fully traceable via an internal spreadsheet
  • Excessive pressure can be fitted manually or automatically
  • Remote maintenance 7 days a week, with remote connection available

Expert advices

  • Calibrated leaks let you carry out regular, easy checks
  • Ensure that there is enough space above the contents for you to be able to take measurements, or use our specially designed needles and accessories
  • Begin the measurement cycle before introducing the needle into the packaging (in order to measure the leak)
  • Air adjustment requires the full attention and skill of a supervisor, given the various environmental variables
  • We would advise connecting to a crimped power supply.
  • We would recommend optimising the measurement speed depending on the target you wish to reach
  • We would suggest regularly saving the data you have collected and making sure your power supplies are safe.
  • Safety: For user protection, we recommend using the sheath incorporated into the appliance.

News video

Watch the video of the measuring instrument EXOS. We explain the operation of this analyzer and the benefits.

Product informations

Ingress protection of enclosureIP 54
Measurement deviceBench-mounted
EnclosureAluminium and Stainless steel
Weight5 kg
Width340 mm
Height170 mm
Depth230 mm
Power100-240VAC -50/60Hz - 63W
Feed pressure4 to 8 Bar (dry and oil free filtred to 20 µm) filtered at 20μm
Inlet pneumatic connection diameter4 x 6 mm tube connection
Memory capacity4 to 32go
DisplayColor touchscreen 7’’
Digital interfaces1 USB 2.0 and 1 Ethernet RJ45
Storage conditions0 - 70°C/0 - 90% HR
Terms of use0 - 40°C/10 - 90% HR
O2 Oxygen measurementTechnology : Electrochemical
Precision : +/- 1% relative 
T90 response time : < 1 s
Resolution : 0.01%
Average service life : 18 months in air
CO2 Carbon Dioxide measurementTechnology : NDIR
Accuracy : +/- 2% full scale
Resolution : 0,1%
T90 response time : 7s
Average service life : 10 years
Pressure measurementyes
Leak measurementTechnology : Mass flowmeter
Precision : +/- 0.8% full scale (μm) ; +/- 1% of the reading + 0.1 ml/min
Resolution : 0.1ml/min - 1μm
Measurement : 0.1-500 ml/min - 15 μm to 500 μm (from 5 μm with the lab option)
Test pressure : From 10 to 500 mbar +/- 1mbar
Flow : Filling: 4200 ml/min
Compressed air supply : 4 to 8 Bar (dry and oil free filtred to 20 µm)
Analytical curves (displayed and recordable)
Burst testTechnology : Barometric
Accuracy : +/- 0,5% full scale
Resolution : 1mbar
Flow scales : 0,2 – 1,4 bar
Quick inflating flow : Minimum 250l/h
Burst pressure : Up to 1,4 bar
Analytical curves (displayed and recordable)
or
Creep test
Technology : Barometric
Accuracy : +/- 0,5% full scale
Resolution : 1mbar
Measurement : 200 - 500 mbar
Pressure : 500 mbar
Flow : Filling: 4200 ml/min
Analytical curves (displayed and recordable)
Blocked circuit detectionyes
Pump suction controlyes
Compensation for atmospheric pressure variationsyes
Results storageyes
USB Connectionyes
Touchscreenyes
TraceabilityFormat, line, operator, batch number, comment at the end of the test
Remote ConnectionRemote control possible for remote maintenance
Network connection (Ethernet socket)yes
Automatic measurement triggeryes
Options3G router + HMI Advanced
Waterproof RJ45 Ethernet port
  • 2.5 L buffer tank with check valve
  • Calibrated leak
  • USB - 1D barcode reader
  • Filter for the expansion valve and oil
  • separator of the compressed air supply
  • Detectable stylus for the touchscreen
  • Can Piercer to pierce metal or rigid membrane packaging
  • Secured needle holder
  • Needle arm support
  • Needle cleaning tool
  • 1 Carry case IP67 : 566x438x216 mm
  • 1 Sampling kit
  • 2 Needles
  • 2 Filters
  • 1 Stabilization tank
  • 1 000 Septums leak
  • 1 Flexible probe holder
  • 1 Metal detected stylus
  • 1 Pressure regulator compressed air
  • 1 Power supply
  • 1 Calibration certificate
  • 1 User manual
Back to top